ANDATECH has adopted a 12-month calibration cycle, replacing the previous 6-month calibration, based on comprehensive testing of our latest fuel-cell sensor technology. 


Our NATA-accredited laboratory has verified that under typical normal personal and workplace conditions, breathalyser sensors remain accurate and stable for up to 12 months.


Today’s Andatech enhanced design sensors are more advanced, more resistant to drift, and designed to maintain long-term accuracy. These improvements enable us to confidently extend the calibration intervals without affecting reliability.


This change also minimises customer downtime and lowers maintenance costs, resulting in a more efficient service experience.


NOTE:

For high-usage or safety-critical environments (e.g., daily workplace testing), a 6-month calibration interval remains advisable to maintain peak performance.